Blood Pressure Units
A blood pressure unit is used to determine the pressure inside the blood vessels of a patient. There are two numbers that are important when taking a patient’s blood pressure, they are the systolic and diastolic pressures. The systolic pressure is higher than the diastolic in a healthy adult and it corresponds to when the heart contracts. While the diastolic number corresponds to when the heart is relaxed between beats. Blood pressure readings are read as a fraction in a way for example 120/80 or 120 over 80 and readings can be classified as either low, normal, or high. A person with low blood pressure has a blood pressure lower than 90/60 and the only real risk associated with this being that the patient may feel light headed or faint if it gets too low. Normal blood pressure is any reading that is higher than 90/60 and lower than 140/90. High blood pressure is any reading higher than 140/90 there are many health risks associated with high blood pressure and any patient with a blood pressure on the high side should consider changing their lifestyle.
How it works
There are two different types of blood pressure units either an automatic or a manual but they both work on the same principle. First, the cuff is wrapped around the patients arm then the cuff is inflated to a point where it cuts off flow. Next, the pressure in the cuff is slowly deflated until vibrations or Korotkoff sounds are heard. When the sounds are first heard this is the systolic pressure of the patient and when the sounds stop this is the diastolic pressure.
Preventative Maintenance
The preventative maintenance of a blood pressure unit consists of, if applicable, an electrical safety inspection followed by a physical inspection of the unit and all accessories. Next the blood pressure unit is tested for accuracy at different blood pressures and the system is tested for leaks. Finally, the battery is checked to ensure that it has been changed within the past 2 years. This is done for 2 reasons the first being so the battery does not corrode and damage the inside electronics. The second reason being, so there is a functional battery in case of an emergency.
Common Errors
Electrical safety – Electrical safety failures are failures found during electrical safety testing and can potentially pose a risk to the patient. This may be a high leakage current, ground wire resistance or too high of an operating current.
Damaged cuff or air hose – A damaged blood pressure cuff or air hose is common especially in a device that is highly used and this is because the Velcro gets worn to a point where it will not hold properly. Blood pressure units also develop leaks over time in the cuff itself or in any of the connectors used to interface the cuff with the air hose. These failures can result in an error or false readings.
Internal Malfunctions – An internal malfunction of a blood pressure unit typically is one of 3 things. First, the motor on the unit may fail; this will cause the cuff to not inflate fast enough or at all. Second, is the pressure transducer which is responsible for reading the pressure inside the circuitry and detecting the vibrations of the patient’s blood pressure. The final internal error that can occur is an internal leak which may give the same errors and false readings as a leaky cuff.
Calibration failure – If a unit will not calibrate when attempted it can be caused by a main board failure, pressure transducer failure or because of a leak in the system. When any of these typically the unit will need to be further assessed to determine exactly what is the cause.
How to get an accurate measurement
- Allow the patient to rest for 5 minutes or more with their arm at heart level and their feet flat on the floor.
- Find the right cuff for the patient as there are multiple sizes for patients with different size arms (If the patient fits in 2 cuff sizes pick the smaller of the 2).
- Wrap the cuff around the arm with the artery marker in line with the patient’s brachial artery so you can comfortably fit one finger under the cuff.
- Inflate the cuff and instruct the patient to no talk or make any sudden movements as this can mimic Korotkoff sounds and give false reading.
- If more even more accurate readings are desired repeat the process on both arms and average the total.
The TRH Value
The added valve of choosing TRH Services to complete your maintenance and repairs on your blood pressure units is that we will ensure that your device will be in great working condition for you and your patients for the next year and it will be fully certified. At TRH Services our expertise can help educate you on your blood pressure systems and make sure that your unit is in top performance. We can provide onsite service to your facility if you are located in South-Western Ontario (London, Toronto, Kitchener, Waterloo, Cambridge, Guelph, Hamilton, etc.).
Common Blood Pressure Units
Listed below is a short list of some common blood pressure units that we work on, a regular basses:
Manufacturer | Model Number |
A&D Engineering Inc. | UA-779 |
A&D Engineering Inc. | UB-521 |
A&D Engineering Inc. | UA-787 |
A&D Engineering Inc. | UA-774 |
A&D Engineering Inc. | UA-767PC |
A&D Engineering Inc. | UA-767 Plus |
A&D Engineering Inc. | UA-767 |
A&D Engineering Inc. | UA-705 |
A&D Engineering Inc. | UA-704 |
A&D Engineering Inc. | UA-702 |
A&D Engineering Inc. | UA-631V |
A&D Engineering Inc. | UA-1020CN |
A&D Engineering Inc. | 5534 |
A&D Engineering Inc. | UA-789 |
Actus | Hand Held |
Almedic | Nova |
Almedic | Hand Held |
Almedica Ag | Hand Held |
ALPK2 | Hand Held |
ALPK2 | DS-125D |
Amber Spa | N/A |
American Diagnostic Corp. (ADC) | 6013 |
American Diagnostic Corp. (ADC) | 6023 |
American Diagnostic Corp. (ADC) | Hand Held |
American Diagnostic Corp. (ADC) | 6021 |
AMG Medical Inc | Professional Series |
AMG Medical Inc | Hand Held |
AMG Medical Inc | 106-960 |
AMG Medical Inc | 106-955 |
AMG Medical Inc | 106-950 |
AMG Medical Inc | 106-945 |
AMG Medical Inc | 106-940 |
AMG Medical Inc | 106-920 |
Barrington Diagnostics | Hand Held |
Barrington Diagnostics | Floor Model |
BMS | 12-525 |
BMS | Hand Held |
Body Check | Hand Held |
BpTRU Medical Devices | BPM-200 |
Briggs Healthcare | 04-620-001 |
BV Medical Standard | Hand Held |
Chattanooga Group | Pressure Feedback |
CK | CK-101 |
E.R.P LTD. | MS-001 |
Healthmate | Hand Held |
Heine Germany | Wall |
Heine Germany | Hand Held |
Heine Germany | Gamma G7 |
Heine Germany | Gamma XXL Wall Mount |
Heine Germany | Gamma G5 |
Heine Germany | Gamma XXL LF |
Henry Schein | Hand Held |
HoMedics, LLC. | BPA-110-2CA |
HoMedics, LLC. | BP-A06-00CA |
IMCO | Hand Held |
Keller Medical Specialties | KMS890 |
Keller Medical Specialties | KMS 890 |
Klinik | Erkameter |
LabTron | Hand Held |
Life Brand | 5534 |
Mabis Health Care Inc. | Hand Held Sphygmomanometer |
Mabis Health Care Inc. | Hand Held |
Mabis Health Care Inc. | 04-244-001 |
Mabis Match Mates | Hand Held |
Mansfield Medical Distributors Ltd | Hand Held |
McCoy Rigby | Single Tube |
Medisana AG | ETP |
Mercurial | Mercury |
Microlife Corporation | BP3MX1-1 |
N/A | Floor Model |
N/A | Hand Held |
Nasco Iso Plastics | LF03956 |
Omron Healthcare Inc. | 12-605 |
Omron Healthcare Inc. | BP760CAN (HEM-7220-CA) |
Omron Healthcare Inc. | HEM-629 |
Omron Healthcare Inc. | HEM-605 |
Omron Healthcare Inc. | HEM-601 |
Omron Healthcare Inc. | HEM-431 |
Omron Healthcare Inc. | HEM-412CRELN2 |
Omron Healthcare Inc. | HEM – 7411CAN |
Omron Healthcare Inc. | Hand Held |
Omron Healthcare Inc. | BP785CAN |
Omron Healthcare Inc. | BP785 (HEM-7222-Z) |
Omron Healthcare Inc. | BP762CAN |
Omron Healthcare Inc. | BP760CAN |
Omron Healthcare Inc. | BP755CAN-CD6 |
Omron Healthcare Inc. | BP742CAN (HEM-7200-CA) |
Omron Healthcare Inc. | BP710CAN |
Omron Healthcare Inc. | BP710(HEM-7113-Z) |
Omron Healthcare Inc. | 964804 |
Omron Healthcare Inc. | 705IT |
Omron Healthcare Inc. | 412CREL |
Omron Healthcare Inc. | 2Z01998L |
Omron Healthcare Inc. | BP785 |
Omron Healthcare Inc. | HEM-7000-CANCS |
Omron Healthcare Inc. | HEM-650 |
Omron Healthcare Inc. | HEM-712c |
Omron Healthcare Inc. | HEM-780 |
Omron Healthcare Inc. | HEM-775CAN |
Omron Healthcare Inc. | HEM-775 CAN |
Omron Healthcare Inc. | HEM-773 |
Omron Healthcare Inc. | HEM-757 |
Omron Healthcare Inc. | HEM-741CAN |
Omron Healthcare Inc. | HEM-741 |
Omron Healthcare Inc. | HEM-7220-CACS |
Omron Healthcare Inc. | HEM-780CANN |
Omron Healthcare Inc. | HEM-712CN2 |
Omron Healthcare Inc. | HEM-780N2 |
Omron Healthcare Inc. | HEM711LXCAN |
Omron Healthcare Inc. | HEM-711DLXCAN |
Omron Healthcare Inc. | HEM-711DLX |
Omron Healthcare Inc. | HEM-711DCXCAN |
Omron Healthcare Inc. | HEM-711ACCANN |
Omron Healthcare Inc. | HEM-711ACCANIN |
Omron Healthcare Inc. | HEM-711 |
Omron Healthcare Inc. | HEM-705CPCAN |
Omron Healthcare Inc. | HEM-705CP |
Omron Healthcare Inc. | HEM-7121-CA |
Omron Healthcare Inc. | HEM-7220-CA |
Omron Healthcare Inc. | HEM-780CAN |
Omron Healthcare Inc. | HEM-650CANN |
Omron Healthcare Inc. | N/A |
Omron Healthcare Inc. | Marshall 85 |
Omron Healthcare Inc. | HEM-907xl |
Omron Healthcare Inc. | HEM-790ITCAN |
Omron Healthcare Inc. | HEM790ITCAN |
Omron Healthcare Inc. | HEM-780N3 |
ParaMed Technology Inc. | 910 |
Prestige Medical | One Pieace |
Prestige Medical | Hand Held |
Rexall Brands Corp | RX800 |
Rexall Brands Corp | RX100 |
Riester GmbH | Wall Mount |
Riester GmbH | Floor Model |
Riester GmbH | Minimus II |
Shoppers Drug Mart, Ltd. | BC03 |
Smiths Medical PM INC. | N/A |
Smiths Medical PM INC. | Mini-Torr Plus |
Sphygmomanometer | Hand Held |
Summit Doppler Systems | Hand Held |
Sunbeam-Oster Company | 7621 |
SunTech Medical Inc. | Hand Held |
Taylor | MOD II |
Taylor | Hand Held |
Technomedic | Hand Held |
Thermor Ltd. | 33601 |
Thermor Ltd. | 3MS1-4K |
Thermor Ltd. | 3AL1-3E |
Tremblay Harrison, Inc. | ABP-A1 |
Tyco Healthcare Group LP | N/A |
Tyco Healthcare Group LP | Wall mounted |
Tyco Healthcare Group LP | Wall Mount |
Tyco Healthcare Group LP | Wall Model |
Tyco Healthcare Group LP | Wall |
Tyco Healthcare Group LP | Super Shock Resistant |
Tyco Healthcare Group LP | One Piece |
Tyco Healthcare Group LP | Mercury Wall Mount |
Tyco Healthcare Group LP | Mercury |
Tyco Healthcare Group LP | Hand Held |
Tyco Healthcare Group LP | Hand |
Tyco Healthcare Group LP | Floor Model |
Tyco Healthcare Group LP | Dual Tube |
Tyco Healthcare Group LP | Cart |
Tyco Healthcare Group LP | Single Tube |
Utility | Hand Held |
Utility | N/A |
Veridian Healthcare LLC | Floor Model |
Vetline | Vet HDO |
VSM MedTech Ltd | BP3-300 |
W. A. Baum Co. Inc. | N/A |
W. A. Baum Co. Inc. | Wall Model |
W. A. Baum Co. Inc. | Wall Mount |
W. A. Baum Co. Inc. | Wall |
W. A. Baum Co. Inc. | Standby |
W. A. Baum Co. Inc. | Riester |
W. A. Baum Co. Inc. | Mercury Wall Mount |
W. A. Baum Co. Inc. | Mercury |
W. A. Baum Co. Inc. | Hand Held |
W. A. Baum Co. Inc. | Floor Model |
W. A. Baum Co. Inc. | Baumanometer |
W. A. Baum Co. Inc. | 1050 |
W. A. Baum Co. Inc. | 0950NL |
W. A. Baum Co. Inc. | Wall Mounted |
Wal-Mart Canada | 20457 |
Welch Allyn | Wall Mount |
Welch Allyn | N344 |
Welch Allyn | Floor Model |
Welch Allyn | D-72417 |
Welch Allyn | 5300 |
Welch Allyn | Hand Held |
Yamasu Pakistan | Hand Held |
Written By: Greg Kishimoto
Edited By: Tyler Hasenpflug CET
Jeff Heilbraun
Hello, I hope all is going well. I wanted to see if you still support the maintenance of the BpTRU-200 device. I work for a cardiac safety core lab in support of the pharmaceutical R&D arena and have an inventory of BpTRU -200 devices. With the BpTRU company closing I am trying to identify resources for technical support and maintenance.
Thanks in advance for your follow-up
Jeff
Jeff Heilbraun
VP Strategic Development
Bioclinica
+1-301-792-5073 cell phone
http://www.Bioclinica.com
Tyler Hasenpflug
HI Jeff,
If you need more information on this please e-mail me at info@trhservices.ca and I will gladly assist you.